Methods
- Sterilisation removes or kills all viable microorganisms; the target sterility assurance level (SAL) is not more than one non-sterile unit in a million (10⁻⁶).
- Moist heat (autoclaving): 121 °C for 15 minutes is the reference cycle, or 134 °C for 3 minutes for instruments; it is the method of choice for aqueous products in sealed containers.
- Dry heat: 160 °C for 2 hours or 180 °C for 30 minutes for glassware, oils and powders; 250 °C or higher depyrogenates glassware.
- Ionising radiation: typically 25 kGy for single-use devices and some drugs; electron beams and X-rays are alternatives to cobalt-60 gamma rays.
- Ethylene oxide gas for heat-sensitive devices; hydrogen peroxide vapour and gas plasma for isolators and instruments.
- Filtration through 0.22 µm sterilising-grade membranes for heat-sensitive solutions, followed by aseptic filling.