Microbiology · Short notes

Sterilization: Short Notes & Cheat Sheet

Sterilisation versus disinfection, heat, filtration, radiation and gas methods, biological indicators, D and Z values, disinfectants and the Spaulding classification.

Short notes

Read these before you attempt the topic's MCQs.

Resistance to killing, from hardest to easiest

decreasing resistancePrionsNeed specialproceduresBacterialsporesSterilisationneededMycobacteriaHigh-leveldisinfectionNon-envelopedvirusese.g. polio,hepatitis AFungi,vegetativebacteriaIntermediate/lowlevelEnvelopedvirusesEasiest: HIV,influenza
Choose the process for the most resistant organism likely to be present. Only sterilisation reliably kills spores.

Definitions

  • Sterilisation kills or removes all microorganisms, including bacterial spores.
  • Disinfection reduces pathogens on inanimate objects but may not kill spores. Antisepsis does the same on living tissue.
  • D value: time at a given temperature to kill 90% of the organisms (one log reduction). Z value: the temperature change that alters the D value tenfold.
  • F₀ is the lethality of a heat process expressed as equivalent minutes at 121 °C. The target sterility assurance level is 10⁻⁶.

Methods

  • Moist heat (autoclave): 121 °C for 15 minutes at about 15 psi, or 134 °C for about 3 minutes. It is the most reliable method, for aqueous solutions, dressings and instruments.
  • Dry heat (hot-air oven): 160 °C for 2 hours (or 170 °C for 1 hour, 180 °C for 30 minutes). Used for glassware, oils and powders that steam cannot penetrate.
  • Filtration through 0.22 µm membranes suits heat-sensitive solutions, but does not reliably remove viruses or mycoplasma. HEPA filters clean air.
  • Gamma radiation (cobalt-60) is 'cold sterilisation' for single-use plastics and devices. UV light only disinfects surfaces and air, because it penetrates poorly.
  • Ethylene oxide gas sterilises heat-sensitive devices and plastics, but it is toxic, carcinogenic and flammable, and items need aeration afterwards.
  • Pasteurisation (63 °C for 30 minutes or 72 °C for 15 seconds) and tyndallisation are not full sterilisation methods.

Disinfectants

  • Alcohol works best at 60–90% (70% is typical), because water is needed to denature proteins. It does not kill spores.
  • Glutaraldehyde 2% is a high-level disinfectant for endoscopes; with long contact it is sporicidal.
  • Chlorhexidine is a skin antiseptic and mouthwash. Quaternary ammonium compounds (benzalkonium) are inactivated by soap.
  • Hypochlorite (bleach) is cheap, broad-spectrum and used for spills and water treatment. Phenol is the reference standard in the Rideal–Walker and Chick–Martin tests.
  • Spaulding classification: critical items (enter sterile tissue) must be sterilised; semi-critical items (touch mucous membranes) need high-level disinfection; non-critical items (touch intact skin) need low-level disinfection.

Cheat sheet

High-yield facts in one place. Cover the right-hand column and test yourself.

Method → conditions → indicator

MethodConditions · biological indicator
Autoclave121 °C, 15 min, 15 psi · Geobacillus stearothermophilus
Hot-air oven160 °C, 2 h · Bacillus atrophaeus
Ethylene oxideGas, 30–60 °C · Bacillus atrophaeus
Gamma radiationCobalt-60 · Bacillus pumilus
Filtration0.22 µm membrane · Brevundimonas diminuta

Spaulding classification

ItemProcess
Critical (surgical instruments)Sterilisation
Semi-critical (endoscopes)High-level disinfection
Non-critical (BP cuff)Low-level disinfection

Memory aids

121 for 15 at 15
Autoclave at 121 °C for 15 minutes at 15 psi.

Exam traps

  • 70% alcohol kills better than absolute alcohol.
  • Dry heat needs a higher temperature and longer time than moist heat.
  • UV radiation disinfects surfaces; it is not a reliable method for sterilising products.
  • A 0.22 µm filter does not reliably remove viruses.