Validation parameters (ICH Q2)
- Specificity: the method measures the analyte without interference from impurities, degradants or excipients.
- Linearity: results are proportional to concentration, usually over at least 5 levels (r often ≥ 0.999).
- Range: the interval with acceptable linearity, accuracy and precision; for an assay usually 80–120% of the test concentration.
- Accuracy: closeness to the true value, shown by recovery (at least 9 determinations over 3 levels; often 98–102%).
- Precision: closeness of repeated results. Repeatability (same analyst, same day), intermediate precision (different days, analysts, equipment) and reproducibility (between laboratories).
- Detection limit (LOD = 3.3σ/S; signal-to-noise about 3:1) and quantitation limit (LOQ = 10σ/S; about 10:1). Robustness: resistance to small deliberate changes (pH, flow rate, temperature).