Pharmaceutical Analysis · Short notes

Spectroscopy: Short Notes & Cheat Sheet

The electromagnetic spectrum, Beer–Lambert law, UV–visible, IR, fluorescence, atomic absorption, NMR and mass spectrometry.

Short notes

Read these before you attempt the topic's MCQs.

The electromagnetic spectrum used in analysis

increasing wavelength, lower energyX-ray< 10 nmCrystal structure(XRD)UV200–400 nmElectronic transitionsVisible400–800 nmColorimetryInfrared4000–400 cm⁻¹Bond vibrationsRadioMHzNMR: nuclear spin
Each region interacts with a different part of the molecule: electrons (UV–visible), bonds (IR) and nuclei (NMR).

UV–visible spectroscopy

  • UV is 200–400 nm and visible is 400–800 nm. Absorption promotes electrons to higher energy levels.
  • Beer–Lambert law: A = εbc, where A is absorbance, ε molar absorptivity, b path length and c concentration. A = log(I₀/I) = −log T.
  • Deviations occur at high concentrations, with stray light, with polychromatic light and when the analyte reacts chemically.
  • Chromophores absorb light (C=C, C=O, aromatic rings); auxochromes (–OH, –NH₂) shift or intensify absorption.
  • Shifts: bathochromic (red, to longer wavelength), hypsochromic (blue, to shorter), hyperchromic (more intense), hypochromic (less intense).
  • Instrument: deuterium lamp (UV) and tungsten–halogen lamp (visible), a grating monochromator, quartz cells for UV (glass absorbs UV), and a photomultiplier or diode-array detector.

IR and fluorescence

  • IR (4000–400 cm⁻¹) measures molecular vibrations. A vibration is IR-active only if it changes the dipole moment.
  • Samples are prepared as a KBr disc, a Nujol mull or by ATR. The fingerprint region below 1500 cm⁻¹ confirms identity.
  • Fluorimetry measures light emitted at a longer wavelength than that absorbed (Stokes shift). It is 10 to 1000 times more sensitive than UV, but suffers from quenching.
  • Fluorescence is measured at 90° to the exciting beam; quinine sulfate is a common standard.

Atomic and structural methods

  • Atomic absorption spectroscopy measures metals using a hollow-cathode lamp made of the element being measured, with a flame or graphite furnace.
  • Flame photometry (emission) measures sodium, potassium, lithium and calcium.
  • NMR uses radio waves on nuclei with spin (¹H, ¹³C). Chemical shift is in ppm relative to TMS (0 ppm). Signals split by the n + 1 rule.
  • Mass spectrometry separates ions by mass-to-charge ratio (m/z). The molecular ion gives the molecular mass; the base peak is the tallest peak (100%).
  • Nitrogen rule: an odd molecular mass means an odd number of nitrogen atoms.

Cheat sheet

High-yield facts in one place. Cover the right-hand column and test yourself.

Technique → what it measures

TechniqueBasis · use
UV–visibleElectronic transitions · assays
IRVibrations · identification
FluorimetryEmission · trace analysis
AASAtomic absorption · metals
Flame photometryAtomic emission · Na, K, Li
NMRNuclear spin · structure
Mass spectrometrym/z · molecular mass, structure

Components and materials

PartChoice
UV sourceDeuterium lamp
Visible sourceTungsten–halogen lamp
UV cellQuartz (silica)
IR sampleKBr disc, Nujol mull, ATR
AAS sourceHollow-cathode lamp
NMR referenceTetramethylsilane (0 ppm)

Memory aids

Bathochromic = Big wavelength (red)
A bathochromic shift moves absorption to longer wavelength; hypsochromic moves it shorter.

Exam traps

  • Glass cells absorb UV light; quartz must be used below about 340 nm.
  • Symmetrical vibrations that do not change dipole moment are IR-inactive.
  • Fluorescence is emitted at a longer wavelength, not a shorter one.
  • The base peak is the most intense peak, not necessarily the molecular ion.

Revised? Now test yourself.

Attempt the Spectroscopy MCQs while the notes are fresh. Anything you get wrong is saved to your incorrect list for the next revision round.